CHARACTERIZE THE ELF: THE NOVEL ELECTRON FLASH IRRADIATION SYSTEM UNVEILED WITH STANDARD DOSIMETRIC TOOLS_

Session Type
FLASH Modalities Track (Oral Presentations)
Date
Wed, 01.12.2021
Session Time
10:20 - 11:30
Room
Room 2.31
Lecture Time
10:40 - 10:50

Abstract

Background and Aims

As the development of suitable detectors in ultra high Dose Per Pulse (DPP) beams is on-going, temporary protocols solving limitations with standard tools are required for the acceptance of new dedicated machines. We present a protocol for system commissioning using clinically available dosimeters, optimizing the use of the detectors. Extra attention was given to beam characterization when varying the DPP.

Methods

Measurements were performed on the ELF, a dedicated research linac for Flash radiotherapy with electrons (ElectronFlash, SIT, Italy) at 7 and 9 MeV using standard dosimetric tools: a PTW Advanced Markus (AM) plane-parallel ion chamber, Gafchromic EBT-XD films, and alanine pellets. The system is highly tunable for pulse length, pulse repetition frequency and number of pulses. Collaboration with the developer allowed further tuning. The DPP was varied using pulse length, distance from the linac and e-gun current.

Results

Performance tests assessing dose, dose stability, PDD, profiles, and output linearity to system settings were measured cross-referencing at least two detectors (comparable within 5%). Energy and dose outputs, monitored with films and AM, were stable within 5% in several months. The AM was used for daily QA for doses up to 0.5 Gy/pulse (model proposed in abstract #211). Changing the DPP using the above-mentioned parameters resulted in a linear output change and minimal variation (<1 MeV) of the spectrum.

Conclusions

We characterized the ELF using standard dosimetric tools. We propose the combination of alanine and Gafchromic films as absolute dosimetric standard. The AM can be employed as real-time tool for daily QA.

Hide

FASTER THAN LIGHT: CAN SCINTILLATORS GUIDE ELECTRON FLASH EXPERIMENTS?

Session Type
FLASH Modalities Track (Oral Presentations)
Date
Wed, 01.12.2021
Session Time
10:20 - 11:30
Room
Room 2.31
Lecture Time
10:50 - 11:00

Abstract

Background and Aims

The radiobiological study of the FLASH effect requires accurate and reliable dosimetry. As scintillators are promising candidates, this work presents a first characterization of their response at UHDR, suggesting solutions to account for possible saturating effects.

Methods

Five scintillating fibers, one clinical and two experimental Y2O3:Eu-based scintillators (DoseVue N.V., Belgium) and two experimental Al2O3:C-based scintillators (SCK CEN, Belgium), were irradiated using the ElectronFlash system (SIT, Italy). Linearity with dose was validated by varying the number of pulses for both experimental Y2O3:Eu-based scintillators. Dose per pulse (DPP) linearity was investigated in all scintillators by varying pulse duration and the distance from the linac exit window (SSD). Pulse scheme stability was investigated by changing of the pulse repetition frequency (PRF).

Results

Good linearity with dose (R²>0.99) was observed in both experimental Y2O3:Eu-based scintillators. Linearity with increasing pulse duration (R²>0.95) and an inverse squared relation between DPP and SSD (R²>0.95) were observed in 4 out of 5 scintillators. However, a concave curvature in response, suggesting saturation, was observed for all scintillators. This effect was more pronounced for the smaller applicator diameter. Our results showed reduced saturation effects when increasing integration time as well as when reducing signal intensity. All scintillators showed a decreasing response with increasing PRF.

Conclusions

The promising characteristics of scintillators as on-line dosimeters for UHDR were validated and possible solutions to reduce saturation effects have been evaluated. Further research on the response by varying PRF is needed.

This work is part of the 18HLT04 UHDpulse project which received funding from the EMPIR programme.

Hide

A FEASIBILITY STUDY OF IORT-FLASH USING A GPU-BASED FAST MONTE CARLO

Session Name
Session Type
FLASH Mechanisms Track (Oral Presentations)
Date
Thu, 02.12.2021
Session Time
11:00 - 12:00
Room
Room 2.15
Lecture Time
11:00 - 11:10

Abstract

Background and Aims

Intra Operative Radiation Therapy (IORT) may represent one of the first clinical modalities of a Flash clinical treatment. Within IORT, whenever needed and possible, temporarily beam modifiers (such as the protection disc for breast carcinoma treatment) are used to protect the underlying healthy tissues during the irradiation. In this contribution we investigate the efficiency achievable in IORT-FLASH treatment using a GPU-based fast Monte Carlo called FRED (Fast particle thErapy Dose evaluator), as a tool for dose calculation and treatment optimization.

Methods

The FRED MC has been developed to allow a fast optimization of the Treatment Planning System in Particle Therapy (simulation time reduced by a factor of 1000), while keeping the dose release accuracy typical of a MC tool. We have simulated in detail the geometry and the material of the applicator coupled with the linac, provided by the SIT company (Aprilia, Italy). We have then combined the FRED simulation with a simple modelling of the FLASH effect and compared it with a conventional IORT treatment.

Results

The tumour coverage and the dose absorbed by the organs at risk have been compared, carrying out a quantitative analysis comparing the obtained Dose Volume Histograms, with a standard IORT treatment.

Conclusions

The results demonstrate the potential of FLASH effect in IORT and of FRED as a tool for treatment planning and dose-report calculations.

Hide

REALIZATION AND CHARACTERIZATION OF NOVEL DIAMOND DETECTOR PROTOTYPES FOR FLASH THERAPY APPLICATIONS

Session Type
FLASH in the Clinic Track (Oral Presentations)
Date
Thu, 02.12.2021
Session Time
11:00 - 12:00
Room
Room 2.31
Lecture Time
11:00 - 11:10

Abstract

Background and Aims

Commercially available real time response dosimetric systems were shown to be unsuitable for Flash therapy, in which ultra-high Dose Per Pulse (DPP) is of interest. There is a need for novel, reliable and fast response detectors for such an application. In the present work, a comprehensive investigation is reported on the properties of diamond-based detectors specifically designed for Flash therapy.

Methods

Several diamond Schottky diode prototypes were produced at Rome Tor Vergata University, in cooperation with PTW Freiburg. Quite a few relevant design and electronical parameters were systematically varied, tuning the device’s overall performance and meet the stringent requirements of ultra-high DPP irradiation. The resulting prototypes were tested by using an ElectronFllsh linac (SIT S.P.A.), by using a 9 MeV electron beam and DPPs up to 12.5 Gy/pulse.

Results

The performance of the prototypes, investigated under a wide range of irradiation parameters, was found to depend on their specific structural properties and designs, showing a response linearity up to a DPP of 12.5 Gy/pulse in the best case.

Conclusions

First results of this systematic investigation, clearly demonstrate the feasibility of a diamond based detector for Flash therapy applications. The analysis of the obtained experimental results will be used as an input for further development and optimization procedure of the investigated prototypes.

The present work is part of the 18HLT04 UHDpulse project which has received funding from the EMPIR programme co-financed by the Participating States and from the European Union’s Horizon 2020 research and innovation programme.

Hide

A NOVEL METHOD FOR DETERMINING IC SATURATION FACTOR (UP TO 0.5 GY/P FOR ADV. MARKUS)

Session Type
FLASH in the Clinic Track (Oral Presentations)
Date
Thu, 02.12.2021
Session Time
11:00 - 12:00
Room
Room 2.31
Lecture Time
11:40 - 11:50

Abstract

Background and Aims

Ionization chambers (IC) represent the standard for performing the commissioning of medical linacs. Nevertheless, their use in the UHDR range is not currently possible, due to the amount of charge produced by each pulse.

For dose-per-pulse (dpp) above 0.5 cGy/p, the approach implemented by international protocols for modelling ion recombination failed, because the free electron fraction p contribution must be considered. We modify the approach of Di Martino (2005) in order to obtain p by means of ionometric measurements only.

Methods

According to the proposed model:

equation1.png

where

qcol is the charge collected by IC;

V is the voltage applied to IC;

qgen is the charge generated by the pulse;

A and λ are parameters depending on the IC..

By varying the voltage applied V, such function can be determined versus the unknown parameters (qgen, A and λ ); then, such parameters can be determined by means of the non-linear regression method.

Once all parameters are known, p is calculated and and then ksat is determined as:

equation2.png

being α = A / V .

Results

The method has been adopted for estimating ksat for the Adv. Markus, both with the beam produced by ElectronFlash and by LIAC HWL.The fit provided an agreement better than 1% within IORT range and better than 5% with 0.6 Gy/p.

fig1.png

Conclusions

Once ksat is known, ionometric measurements at the larger distance might become the central element of a reliable Quality Assurance program for any Flash linac.