Ralf-Peter Kapsch (Germany)

Physikalisch-Technische Bundesanstalt (Braunschweig) 6.21

Author Of 3 Presentations

OVERVIEW AND CURRENT STATUS OF THE JOINT RESEARCH PROJECT UHDPULSE - “METROLOGY FOR ADVANCED RADIOTHERAPY USING PARTICLE BEAMS WITH ULTRA-HIGH PULSE DOSE RATES”

Session Type
FLASH Modalities Track (Oral Presentations)
Date
Wed, 01.12.2021
Session Time
14:50 - 15:50
Room
Room 2.15
Lecture Time
14:50 - 15:00

Abstract

Background and Aims

Dosimetry for FLASH radiotherapy, VHEE radiotherapy as well as for laser-driven beams cause significant metrological challenges due to the ultra-high dose rates and pulsed structure of these beams, in particular for real time measurements with active dosimeters. It is not possible to simply apply existing Codes of Practice available for dosimetry in conventional external radiotherapy here. However, reliable standardized dosimetry is necessary for accurate comparisons in radiobiological experiments, to compare the efficacy of these new radiotherapy techniques and to enable safe clinical application. UHDpulse aims to develop the metrological tools needed for reliable real-time absorbed dose measurements of electron and proton beams with ultra-high dose rate, ultra-high dose per pulse or ultra-short pulse duration.

Methods

Within UHDpulse, primary and secondary absorbed dose standards and reference dosimetry methods are developed, the responses of available state-of-the-art detector systems are characterised, novel and custom-built active dosimetric systems and beam monitoring systems are designed, and methods for relative dosimetry and for the characterization of stray radiation are investigated.

Results

Prototypes of different active dosimetry systems show promising results for real-time dosimetry for particle beams with ultra-high pulse dose rates. The results of the UHDpulse project will be the input data for future Codes of Practice.

Conclusions

A brief overview of the progress in the UHDpulse project and the involved institutions will be given.

Acknowledgement: This project 18HLT04 UHDpulse has received funding from the EMPIR programme co-financed by the Participating States and from the European Union’s Horizon 2020 research and innovation programme.

Hide

ION COLLECTION EFFICIENCY IN ULTRA-HIGH DOSE PER PULSE ELECTRON BEAMS

Session Type
FLASH Modalities Track (Oral Presentations)
Date
Wed, 01.12.2021
Session Time
14:50 - 15:50
Room
Room 2.15
Lecture Time
15:40 - 15:50

Abstract

Background and Aims

Ionometry is challenging in ultra-high dose per pulse (DPP) beams due to high level of ion recombination and it does not follow the current theoretical model. This investigation aims to measure the ion collection efficiency for different types of ionization chambers.

Methods

The response of three plane-parallel ionization chambers has been measured in an ultra-high DPP (0.5 to 2 Gy/pulse) 20 MeV electron beam. To measure the pulse charge, a non-destructive integrating current transformer (ICT) is used. The ICT signal was calibrated against the absorbed dose to water measured with alanine.

Results

The dependence of the ion collection efficiency is not linear with the DPP. The intra-type variations were evaluated to be in the 2-5% range as illustrated in figure 1. The ion collection efficiency of the Advanced Markus chamber has been compared to the Petersson et al. (DOI: 10.1002/mp.12111) empirical model. The difference between the measured and expected values from their model is consistent with the intra-type variation observed.

ionometry.png

Figure 1: Ion collection efficiency in ultra-high dose per pulse electron beams

Conclusions

The investigation has shown that the ion collection efficiency of all ion chambers used for dosimetry in ultra-high DPP beams should be evaluated to improve uncertainty. It is important to develop a new theoretical model for ion recombination in ultra-high DPP beams.

Acknowledgement: This project 18HLT04 UHDpulse has received funding from the EMPIR programme co-financed by the Participating States and from the European Union’s Horizon 2020 research and innovation programme.

Hide

PERFORMANCE OF A PROBE-TYPE GRAPHITE CALORIMETER (AERROW) IN ULTRA-HIGH DOSE PER PULSE ELECTRON BEAMS

Session Type
FLASH in the Clinic Track (Oral Presentations)
Date
Thu, 02.12.2021
Session Time
11:00 - 12:00
Room
Room 2.31
Lecture Time
11:20 - 11:30

Abstract

Background and Aims

One of the challenges in investigating FLASH radiation therapy is the reliable measurement of the absorbed dose at ultra-high dose per pulse (DPP). This work aims to evaluate a probe-type graphite calorimeter (Aerrow) at ultra-high DPP.

Methods

The DPP was varied between 0.5 to 5.5 Gy. The signal was compared to the pulse charge measured by a non-destructive Integrating Current Transformer (ICT). In addition, a depth dose curve was measured and compared to an ionization chamber measurement and a Monte-Carlo simulation.

Results

The response of the calorimeter was proportional to the ICT signal. On average, the standard deviation of the calorimeter response was 0.2 % in high dose rate and reached 1% at the lower rate. In figure 1, the depth dose measurement with calorimeter and ionization chamber is compared to Monte-Carlo simulation. The half-value depths in water R50 obtained from measurement and simulation agree within 1 mm.

calorimeter.png

Figure 1: Depth dose in water measured and simulated.

Conclusions

Calorimetry is showing promising results for absorbed dose measurements. Calorimetry gets simpler at FLASH dose rates as the dose delivery is in a few seconds or less. The preliminary results show that advanced thermal insulation of the calorimeter is not required, nor the use of a heat lost correction factor.

Acknowledgement: This project 18HLT04 UHDpulse has received funding from the EMPIR programme co-financed by the Participating States and from the European Union’s Horizon 2020 research and innovation programme.

Hide