AS01 Closed-loop System and Algorithm

503 - PERFORMANCE OF THE LILLY AUTOMATED INSULIN DELIVERY (AID) SYSTEM: RESULTS OF EARLY PHASE FEASIBILITY STUDY

Session Type
E-POSTER DISCUSSION
Session Name
E-POSTER DISCUSSION 01

Abstract

Background and Aims

AID systems have led to demonstrated improvements in time in range without increasing hypoglycemia. Further advances in performance and usability will set the stage for more widespread adoption of this technology in diabetes self-management.

Methods

This inpatient early feasibility study evaluated the Lilly hybrid closed loop system comprising a high accuracy pump (developed by DEKA Research & Development Corp.), Lispro insulin, pump-embedded model predictive control algorithm, Dexcom continuous glucose monitor, and smartphone-based controller. Subjects underwent two similar 2-day inpatient protocols (10 adults with type 1 diabetes per protocol) including meal-related challenges (~30% over-bolus and 1-hour delayed bolus encompassing carbohydrate coverage and hyperglycemia correction) to simulate real-world diabetes self-management errors.

Results

Subjects (25% male) had mean age: 44.7±14.2 years, T1D duration: 30.2±11.1 years, A1C: 7.2±0.8%, and insulin usage: 0.53±0.21 U/kg/day. The table below describes glucose control (percent time in different ranges [mean±SD]) while the subjects were on the AID system.

table_insu-85877.png

Conclusions

In this closely supervised feasibility study the Lilly AID system demonstrated expected algorithm performance in response to simulated diabetes self-management errors and promising glycemic outcomes. Clinical trial evidence in more ‘real world’ environments will provide additional information on the safety and effectiveness of the system.